Ivoclar Programat P95 Manual May 2026
The Ivoclar Programat P95 is a laboratory furnace used in dental prosthetics for sintering and firing ceramic materials (e.g., zirconia, lithium disilicate, veneering porcelains) and for glazing, crystallization, and staining procedures. Designed for dental labs and clinics, it emphasizes process control, reproducibility, and integration with modern dental workflows.
| Cycle type | Typical temp | Hold time | Vacuum | |------------|--------------|-----------|--------| | e.max Press (200g ring) | 920–940°C | 15–20 min | Yes | | e.max Crystallisation | 840–850°C | 25–30 min | No | | Empress Esthetic | 1070°C | ~10 min | Yes |
A: The service mode is reserved for certified technicians. The manual only provides a warning: “Unauthorized entry may void warranty.” To enter, you typically need a special code provided by Ivoclar technical support. Ivoclar Programat P95 Manual
The manual dedicates its first major section to safe installation. According to the official guidelines:
The P95 diagnostic system outputs specific codes. The manual provides the following table: The Ivoclar Programat P95 is a laboratory furnace
| Error Code | Meaning | Operator Action | | :--- | :--- | :--- | | E-12 | Thermocouple broken (infinite resistance) | Replace type B TC. | | E-23 | Heating rate too slow (element aging) | Run regeneration cycle (1500°C, 2h, no load). | | E-45 | Vacuum leak | Check door seal and hose connections. | | E-67 | Cooling fan stalled | Clean fan blades; check 12V DC supply. | | E-88 | Overtemperature (>1650°C) | Immediate reset. Check SSR relay for short circuit. |
The manual lists all pre-installed programs. Example: The manual only provides a warning: “Unauthorized entry
| Program Number | Material | Type | Start Temp | Hold Time | | --- | --- | --- | --- | --- | | 1 | IPS e.max Press | Pressing | 700°C | 15:00 min | | 5 | IPS e.max CAD | Crystallization | 750°C | 10:00 min | | 12 | IPS InLine | Firing | 580°C | 1:00 min |
Never delete factory program 0 – it contains the service and calibration menu.
Based on the Ivoclar service manual, adhere strictly to this timeline:
| Interval | Action | | :--- | :--- | | Daily | Wipe chamber floor with vacuum aspirator. Clean glass viewing port. | | Weekly | Calibrate thermocouple using the TC1 Test (requires Ivoclar calibration ring). | | Monthly | Inspect MoSi₂ elements for pitting. Rotate sintering platform 180°. | | Quarterly | Run "Service Mode > Element Resistance." Normal range: 1.2 – 2.8 Ohms (cold). Replace if >4.0 Ohms. | | Annually | Factory calibration of vacuum sensor and thermocouple. Replace door gasket. |