Syngistix for AA is a workflow-based software platform designed by PerkinElmer to control atomic absorption (AA) instruments like the PinAAcle series. It is structured around a ribbon-style interface that guides users through five main functional areas to streamline the journey from sample to results. Software Interface & Navigation
The software uses a "left-to-right" workflow represented by five primary tabs on the main ribbon:
Instrument: Controls the spectrometer, atomizer (flame or furnace), and autosampler. It provides real-time system diagnostics and instrument status.
Analysis: Used for managing methods, editing features, and inputting sample information. Users set up analytical runs and batch sequences here.
Results: Dedicated to data review, calibration curve editing, and reprocessing previously acquired data.
Utilities: Provides access to supplemental applications and interface tools.
Techniques: Allows selection of the running technique, such as Flame, Graphite Furnace, or specialized methods like FIMS (Flow Injection Mercury System). Core Procedures 1. Method Setup Summary
Choose Element: Select the target element and its standard conditions (wavelength, slit width, lamp current) from the software library.
Calibration: Define calibration standards and units. Syngistix allows for "Recall Calibration" to reuse stored curves or manual editing to meet statistical requirements. 2. Running an Analysis
Sample Information: Use the Sample Information Editor to define sample IDs and positions.
Initiate Batch: Highlight samples in the Batch Tab and select "Analyze Batch". The software then coordinates the spectrometer and autosampler according to the active method. 3. Data Management & Reprocessing Product: Syngistix for AA Software - PerkinElmer
| Error Message | Probable Cause | Software Fix (from manual) | | :--- | :--- | :--- | | "Lamp Energy Low" | Lamp misaligned or aged | Run the "Lamp Optimization" routine. Check lamp current settings. | | "Flame Out" | Gas pressure drop or leak | Check gas regulators; click "Reset Flame" in the Safety Panel. | | "Furnace Over-temp" | Cooling water flow failure | Halt run; check chiller water flow rate in the System Status tab. | | "Autosampler Position Failed" | Obstructed rack or motor stall | Use "Jog" controls to manually move the arm; re-home the autosampler. |
| Flag | Meaning | |------|---------| | H | High (> high cal std) | | L | Low (< detection limit) | | QC-CCV | Out of recovery | | D | Duplicate RPD fail |
Syngistix for AA is the flagship software platform used to control PerkinElmer atomic absorption spectrometers (such as the PinAAcle and NexION series). It is designed to unify Flame, Graphite Furnace (GFAAS), and ICP-MS workflows into a single interface. Clarity & Organization